PACKAGED RO/DI SYSTEMS

QuickLab laboratory water purification systems.

Compare four QuickLab configurations for RO or RO/DI water. Choose a starting point by water quality, demand, storage and distribution requirements, then review your installation with PPT.

QuickLab equipment with storage tank, controls and treatment vessels installed in a laboratory utility room
Product overview

Engineered as a system. Packaged for your lab.

QuickLab is a standardized family of two production platforms for laboratory water. Pre-engineered RO and RO/DI configurations combine treatment, storage, loop distribution and PLC/HMI controls in a compact package, with sectional breakdown for installation through restricted access.

  • Compact by design

    Sectional breakdown for restricted access.

  • Pre-engineered platforms

    Two base systems. Four configurations.

  • Defined water quality

    RO and RO/DI configurations.

QuickLab configurations

Four configurations.
Two production platforms.

Choose the platform for water demand, then select RO or Plus for water quality. Compare the equipment here and review the published specifications below.

Distribution pump ratings shown at 50 PSI. Final loop performance depends on your installation.

Technical comparison

Compare the configurations.

Swipe across the table to compare all four models.

QuickLab model specifications
QuickLab 275RO waterQuickLab 275 PlusASTM Type III waterQuickLab 450RO waterQuickLab 450 PlusASTM Type II/III water
RO production1.25 GPM1.25 GPM2 GPM2 GPM
Water qualityRO waterASTM Type III waterRO waterASTM Type II/III water
Distribution pump at 50 PSI12 GPM12 GPM20 GPM20 GPM
Storage capacity250 gal250 gal250 gal250 gal
Supply / return loop1 in. (32 mm)1 in. (32 mm)1¼ in. (40 mm)1¼ in. (40 mm)
Installed footprint (W × D × H)8 × 4 × 7 ft10 × 4 × 7 ft8 × 4 × 7 ft12 × 4 × 7 ft
Full specificationsDownload PDFDownload PDFDownload PDFDownload PDF

Footprints are published equipment dimensions; allow additional installation and service clearance. Match the required grade to the selected configuration.

Close view of QuickLab controls, pumps, filter housings and hard piping on an installed system
Detail of an installed QuickLab system.
Engineering & treatment

Treatment and distribution, configured together.

QuickLab combines pretreatment, reverse osmosis and optional DI polishing with integrated controls, storage and distribution. The selected water quality determines the treatment package.

  1. Pretreatment

    Selected for feedwater

  2. Reverse osmosis

    Primary separation

  3. DI polishing

    Plus configurations

  4. UV + final filtration

    Application-selected

  5. Storage + distribution

    Tank, pump and loop

Conceptual sequence, not a piping diagram. DI applies to Plus. Confirm pretreatment, 254 nm UV and final filtration against the selected configuration.

Controls, materials and service access

PLC/HMI controls provide alarms and trends. Model sheets specify SCH80 PVC before RO and polypropylene socket-fusion piping after RO. Confirm facility integration, final connections and access to consumables with PPT.

View technical documents

Applications and planning.

QuickLab supports analytical and research laboratories, clinical diagnostics and small production utilities. Start with water quality and demand, then work through your facility requirements.

Request a Spec Review

For drawings, specifications or defined equipment requirements.

  • Required water quality

    Match the selected model’s water quality to each method, instrument and point of use.

  • Demand, feedwater & utilities

    Review daily use and peak draws, feedwater chemistry, electrical supply and drains.

  • Storage & distribution

    Coordinate 250-gallon storage with loop head, simultaneous draws and turnover.

  • Installation & service access

    Confirm doorways, transport sections, final connections and access to consumables.

Installation, startup and ongoing support

QuickLab ships mostly preassembled, with single-point utility connections and front-access consumables. Plan final connections and commissioning around your room. Review filters, membranes, DI media and UV equipment where fitted, then agree the service scope with PPT.

Water services & support
Feedwater and utility criteria

Published model requirements include 30–60 PSI feedwater pressure and inlet flow of 4 GPM for QuickLab 275, 8 GPM for 275 Plus and 6 GPM for both 450 models. Review the selected model, water analysis and site connections with PPT.

Review model documents

Bring your lab water requirements to PPT.

Have drawings, specifications or a defined equipment schedule? Send them for a Spec Review. For an open application question, talk with the team.

Privacy settings

Choose whether to allow optional analytics. You can change your choice at any time using Privacy Settings in the footer.

Necessary functionality

Supports your privacy choices and features you request, such as inquiry forms and Osmo. Rejecting analytics does not disable these features.

Always available

Read our Privacy Policy and Cookie Policy.