QuickLab laboratory water purification systems.
Compare four QuickLab configurations for RO or RO/DI water. Choose a starting point by water quality, demand, storage and distribution requirements, then review your installation with PPT.

Engineered as a system. Packaged for your lab.
QuickLab is a standardized family of two production platforms for laboratory water. Pre-engineered RO and RO/DI configurations combine treatment, storage, loop distribution and PLC/HMI controls in a compact package, with sectional breakdown for installation through restricted access.
Compact by design
Sectional breakdown for restricted access.
Pre-engineered platforms
Two base systems. Four configurations.
Defined water quality
RO and RO/DI configurations.
Four configurations.
Two production platforms.
Choose the platform for water demand, then select RO or Plus for water quality. Compare the equipment here and review the published specifications below.

QuickLab 275
RO water
- RO production
- 1.25 GPM
- Distribution pump
- 12 GPM
- Storage
- 250 gal

QuickLab 275 Plus
ASTM Type III water
- RO production
- 1.25 GPM
- Distribution pump
- 12 GPM
- Storage
- 250 gal

QuickLab 450
RO water
- RO production
- 2 GPM
- Distribution pump
- 20 GPM
- Storage
- 250 gal

QuickLab 450 Plus
ASTM Type II/III water
- RO production
- 2 GPM
- Distribution pump
- 20 GPM
- Storage
- 250 gal
Distribution pump ratings shown at 50 PSI. Final loop performance depends on your installation.
Compare the configurations.
Swipe across the table to compare all four models.
| QuickLab 275RO water | QuickLab 275 PlusASTM Type III water | QuickLab 450RO water | QuickLab 450 PlusASTM Type II/III water | |
|---|---|---|---|---|
| RO production | 1.25 GPM | 1.25 GPM | 2 GPM | 2 GPM |
| Water quality | RO water | ASTM Type III water | RO water | ASTM Type II/III water |
| Distribution pump at 50 PSI | 12 GPM | 12 GPM | 20 GPM | 20 GPM |
| Storage capacity | 250 gal | 250 gal | 250 gal | 250 gal |
| Supply / return loop | 1 in. (32 mm) | 1 in. (32 mm) | 1¼ in. (40 mm) | 1¼ in. (40 mm) |
| Installed footprint (W × D × H) | 8 × 4 × 7 ft | 10 × 4 × 7 ft | 8 × 4 × 7 ft | 12 × 4 × 7 ft |
| Full specifications | Download PDF | Download PDF | Download PDF | Download PDF |
Footprints are published equipment dimensions; allow additional installation and service clearance. Match the required grade to the selected configuration.

Treatment and distribution, configured together.
QuickLab combines pretreatment, reverse osmosis and optional DI polishing with integrated controls, storage and distribution. The selected water quality determines the treatment package.
Pretreatment
Selected for feedwater
Reverse osmosis
Primary separation
DI polishing
Plus configurations
UV + final filtration
Application-selected
Storage + distribution
Tank, pump and loop
Conceptual sequence, not a piping diagram. DI applies to Plus. Confirm pretreatment, 254 nm UV and final filtration against the selected configuration.
Controls, materials and service access
PLC/HMI controls provide alarms and trends. Model sheets specify SCH80 PVC before RO and polypropylene socket-fusion piping after RO. Confirm facility integration, final connections and access to consumables with PPT.
View technical documentsApplications and planning.
QuickLab supports analytical and research laboratories, clinical diagnostics and small production utilities. Start with water quality and demand, then work through your facility requirements.
Request a Spec ReviewFor drawings, specifications or defined equipment requirements.
Required water quality
Match the selected model’s water quality to each method, instrument and point of use.
Demand, feedwater & utilities
Review daily use and peak draws, feedwater chemistry, electrical supply and drains.
Storage & distribution
Coordinate 250-gallon storage with loop head, simultaneous draws and turnover.
Installation & service access
Confirm doorways, transport sections, final connections and access to consumables.
Installation, startup and ongoing support
QuickLab ships mostly preassembled, with single-point utility connections and front-access consumables. Plan final connections and commissioning around your room. Review filters, membranes, DI media and UV equipment where fitted, then agree the service scope with PPT.
Water services & supportFeedwater and utility criteria
Published model requirements include 30–60 PSI feedwater pressure and inlet flow of 4 GPM for QuickLab 275, 8 GPM for 275 Plus and 6 GPM for both 450 models. Review the selected model, water analysis and site connections with PPT.
Review model documentsBring your lab water requirements to PPT.
Have drawings, specifications or a defined equipment schedule? Send them for a Spec Review. For an open application question, talk with the team.