Product overview

RO water with integrated storage and distribution.

QuickLab 450 combines 2 GPM reverse-osmosis production with a 250 gal storage tank, 20 GPM distribution pump rated at 50 PSI, and PLC/HMI controls. The model sheet lists RO water quality based on the incoming water conditions.

  • Packaged treatment

    RO water for laboratory utilities.

  • Integrated distribution

    Storage, pump and return-loop monitoring.

  • PLC/HMI controls

    System operation and water-quality visibility.

Technical specifications

QuickLab 450 specifications.

System performance, water requirements and installation connections.

Model specification & documents
Water quality
RO water
RO production
2 GPM
Distribution capacity
20 GPM
Pump rated at 50 PSI
Storage capacity
250 gal

Performance & water quality

Recovery rate
70%
RO reject rate
95%–97%
Water quality
10–40 μS/cm (based on inlet <750 μS/cm)
TOC, sodium, chlorides and total silica
Performance based on inlet concentration. Contact PPT for evaluation.

Confirm application-specific water-quality requirements with PPT.

Storage & distribution

Storage tank
250 gal; 0.2μ filter, spray ball and overflow
Distribution pump
20 GPM @ 50 PSI (max 600 ft loop)
Return piping
Flowmeter, BPR and gauge

Pump performance depends on the installed distribution loop.

Feedwater requirements

Hardness
<150 ppm
Total chlorine
0.5 ppm max
Silica
<30 ppm
Temperature
1–30 °C

Utilities & connections

City water flow
6 GPM
City water pressure
30–60 PSI
Power
480 V, 3 phase, 15 A
Inlet connection
¾ in. NPT
Drain connection
1½ in. NPT
Supply / return loop
1¼ in. (40 mm) PPL socket fusion

Allow installation and maintenance clearance beyond the published equipment footprint.

Physical construction & controls

Installed footprint (W × D × H)
8 × 4 × 7 ft
Controls
PLC/HMI; UL-listed panel
Skid frame
Structural aluminum
Piping, monitoring & treatment components
Pre-RO piping
SCH80 PVC
Post-RO piping
PPL socket fusion
UV
254 nm; monitor and intensity
Final filter
Multi-cartridge stainless steel; 0.2μ filter (source notation)
Monitoring
Conductivity at loop out and return
Installed QuickLab family treatment equipment, controls and piping; representative installation detail
Engineering & operation

RO treatment. Monitored distribution.

QuickLab 450 combines treatment with storage, UV, final filtration and loop monitoring. Pretreatment is selected around incoming water quality; the final connections follow the selected specification.

  1. Reverse osmosis

    Primary treatment of preconditioned feedwater.

  2. UV and final filtration

    254 nm UV and final filtration per model sheet.

  3. Storage and distribution

    250 gal tank and 20 GPM distribution pump.

QuickLab family installation detail. Treatment summary is conceptual; final stage order, connections and pretreatment follow the selected model and feedwater requirements.

Plan your installation.

Resolve your water requirements, utilities and access before ordering.

  • Water quality and use

    Confirm your application needs RO water. Review additional polishing separately if your point of use requires a different quality.

  • Demand and distribution

    RO make-up and distribution pump flow are different ratings. Review your daily use and peak draws against 2 GPM RO make-up, 250 gal storage and the distribution pump rating. Plan your loop around your use points.

  • Feedwater and utilities

    Bring your feedwater analysis and your proposed utility connections. Check your water chemistry, power, pressure and drainage against the selected model sheet.

  • Delivery and service access

    Published dimensions describe the equipment. Allow additional installation and service space; review your access route, tank position and clearance for filters, UV and controls.

Service and maintenance

Consumables and sanitization

Plan filters, UV upkeep, sanitization and service intervals around your installed configuration and actual use.

Parts and technical support

Keep your model information and documentation available for service, troubleshooting and replacement parts.

Water Services & Support

Review QuickLab 450 for your lab.

Send your required water quality, daily use, peak draws and feedwater analysis. Include available plans so PPT can review this configuration before procurement.

Privacy settings

Choose whether to allow optional analytics. You can change your choice at any time using Privacy Settings in the footer.

Necessary functionality

Supports your privacy choices and features you request, such as inquiry forms and Osmo. Rejecting analytics does not disable these features.

Always available

Read our Privacy Policy and Cookie Policy.